Strany 1

1Eagle & Eagle X Owners Manual

Strany 2 - EAGLE & EAGLE X OvErviEw

10Air COnDitiOninG / hEAt pUmpbOiLEr & wAtEr hEAtErs & hiGh EffiCiEnCy mODULAtinG hOt wAtEr systEmsfUrnACEs: 80% fUrnACEs: 90%Suction Line:

Strany 3 -  • N/F = Not Found:

11fUrnACEs (COntinUED): AtmOsphEriC, GAs & OiLnAtUrAL GAs & prOpAnEVerify proper combustion: •O2 •CO •StackTemp •VentPressure •Eff

Strany 4 - Getting Started

12OthEr impOrtAnt fACtOrs rELAtinG tO COmbUstiOn •ThethreeT’sofcombustion – Time •Amountoftimethatthefuelandoxygenaretogetherin

Strany 5 - Quick Start Guide

13version has been ignored. This loss is subtracted from the efficiency.CO Air frEE Certain standards ( ANSI Z21.1) for Carbon Monoxide are stated in

Strany 6

14GEnErAL mAintEnAnCE  •Calibrateyourinstrumentannuallytoensureitmeetsoriginalperformancespecifications •Keepyourinstrumentdry.If

Strany 7

15bAttEriEs rEpLACEmEnt This meter has been designed for use with both alkaline  andrechargeableNickelMetalHydride(NiMH)batteries. No other

Strany 8 - Ain mEnU nAviGAtiOn

EAGLE SPECIFICATIONSEAGLE & EAGLE X LimitED wArrAnty The Eagle Combustion Analyzers (C75, C125, C127) are warrantied to be free from defects in m

Strany 9 - GAs firED pOwEr bUrnErs

21.800.547.5740 • WWW.UEiTEST.COMEAGLE & EAGLE X OvErviEwInfrared Printer Port WorklightOn/Off ButtonNavigation “s“UPand BacklightNavigation

Strany 10 - WHERE TO TEST

3DispLAy pArAmEtErs On UEi COmbUstiOn AnALyzErs • TF = Flue Temperature: Calculate net temperature in °F or °C. - Shows ambient temp

Strany 11 - AtUrAL GAs & prOpAnE

4 sAfEty nOtEs Before using this meter, read all safety information carefully. “WARNING” is used to indicate conditions or actions that may pose

Strany 12 - COmbUstiOn mEAsUrEmEnt tErms

5Quick Start Guide .Turn Power on in area of fresh air and allow to countdownRotate test selector to Fuel. Press “s“UP or “t“DOWN to scroll and select

Strany 13 - CO Air frEE

6hiGh rEsOLUtiOn mAnOmEtEr DrAft & stAtiC prEssUrE tEstPress “SEND” to start the printout of results from any test screen.Press “SEND” again duri

Strany 14 - GEnErAL mAintEnAnCE

Quick Start Guide .7Rotate test selector to Exch Test. Call for heat on the system. Observe and wait for O2 readings to stabilize.Prior to the blower

Strany 15 - UsinG rE-ChArGEAbLE bAttEriEs

8Use the “s“UP or “t“DOWN buttons to scroll through the MAIN MENU or OPTIONS. Press “SEND” to select displayed screen choice. NOTE: Rotate the select

Strany 16 - EAGLE SPECIFICATIONS

9AtmOsphEriC GAs firED bUrnErs • Oxygen .....................7to9%O2 • Stack

Komentáře k této Příručce