KM9106 Operators Manual 11. ANALYSER LAYOUT AND FEATURES 1.1 Handset Features Rubber bootRemote lead socket(15 pin D type)Also used fordownloading
KM9106 Operators Manual 10 4.3.2 Line Scroll Mode Line scroll mode allows you to customise the display. Use the key to change the bottom line o
KM9106 Operators Manual 114.4 Sampling the Flue Gas Once the automatic calibration procedure has been completed and the specific fuel has been select
KM9106 Operators Manual 12 4.5 Taking a Pressure Reading With the optional pressure module fitted a flue draught measurement can be made at any time.
KM9106 Operators Manual 134.7 Normal Shutdown Sequence DO THIS EVERY TIME YOU USE THE ANALYSER. Remove the probe from the flue - TAKE CARE ! THE PRO
KM9106 Operators Manual 14 5. USING THE KEYPAD 5.1 Basic Operation Basic operation of the keypad to change the display in PAGE and LINE SCROLL mode
KM9106 Operators Manual 15UPPER PARAMETER KEYS ALWAYS Press UPPER FUNCTION first then SCALE to display Scaling setup paramete
KM9106 Operators Manual 16 5.4 DISPLAY BACKLIGHT AND CONTRAST The display contrast can be adjusted to suit different lighting conditions and the bac
KM9106 Operators Manual 176. USER SELECTABLE SETTING The following features are under your control at any time and can be changed as detailed later i
KM9106 Operators Manual 18 they drift. This may happen if the sensor is taken to very high concentrations for long periods of time or over-ranged.
KM9106 Operators Manual 19To Set The Time: TIME 16-01-12 SET As with the date each number in the time is to be set. Parameter Controls Op
KM9106 Operators Manual 2 1.2 Analyser Layout Connection leadfor OxygensensorOxygen Sensor Filter Bridgeor Sulphur FilterParticle FilterBatteryenclos
KM9106 Operators Manual 20 To Set A Standard Fuel: There are over 70 standard fuels programmed into Quintox. The fuels are arranged into tables for e
KM9106 Operators Manual 21To Set A User Fuel: If one of the standard fuels does not approximate to the one you are using in your boiler then it is po
KM9106 Operators Manual 22 TO SET GROSS OR NETT EFFICIENCY Press the -- -- -- -- -- -- -- -- -- -- -- -- -- SET -- -- -- -- -- -- -- -- -- -- --
KM9106 Operators Manual 23 Parameter Controls OptionsSELECT LANGUAGE Select using or key and ENGLISHSPANISHNETHERLAN
KM9106 Operators Manual 24 Parameter Controls OptionsCOMPENSATION Select using or key and OFFON SET TEMPERATURE CELSIUS
KM9106 Operators Manual 25Selecting YES allows you to set an oxygen reference value different to that shown in the fuel constants. For example 3% is s
KM9106 Operators Manual 26 Parameter Controls OptionsREF NOx Select using or key and NOx = SUMNOx = NONOx = NO2 There are
KM9106 Operators Manual 276.5 Inlet Temperature The Quintox uses as default the internal ambient sensor when calculating the Net temperature. If an
KM9106 Operators Manual 28 6.6 Oxygen Calibration If used over long periods the Oxygen sensor may drift slightly and for maximum accuracy may requi
KM9106 Operators Manual 29 CALIBRATE SENSOR Select using or key and YESNO Select YES SET ZERO ? NO Parameter Controls Opti
KM9106 Operators Manual 31.3 Standard Probe Configuration
KM9106 Operators Manual 30 6.8 CO Alarm It is possible to set a point in the range of the sensor so that it alarms and warns the user of a dangerou
KM9106 Operators Manual 317. PRINTING INFORMATION While in any of the main displays a manual print can be obtained by pressing The display will show
KM9106 Operators Manual 32 The remainder of this section explains the following :- Setting auto-timed printing or logging Allows information to be
KM9106 Operators Manual 33The next screen enables or disables the integral printer. Default is ON. PRINTER ? ON Use or key to selec
KM9106 Operators Manual 34 7.2 User Defined Printouts General Principle: A user defined printout can have a maximum of 40 lines. The contents of ea
KM9106 Operators Manual 35The above example will send the Time and Oxygen reading to the printout every two minutes. To stop the Quintox Auto printin
KM9106 Operators Manual 36 7.2.2 Master List of Printed Parameters The following list is a master list of printed parameters. Details of the measured
KM9106 Operators Manual 378. QUINTOX LOGGING AND PC DOWNLOAD 8.1 Overview 8.1.1 Description Data is sent to the handset through the connection lead
KM9106 Operators Manual 38 8.1.3 Logging Data Data logging is done with the handset operating in 2. QUINTOX CONTROL and by either pressing the
KM9106 Operators Manual 398.3 Logger Control This facility allows access to all of the information stored in the handset; data can be cleared, viewed
KM9106 Operators Manual 4 1.4 Analyser Connections
KM9106 Operators Manual 40 LOG DATA ? NO ENTER YES to tell the logger to store data in memory and NO to disable the logging function. DOWNLO
KM9106 Operators Manual 41ENTER the desired location to display the memory FROM. Entering numbers is detailed in section 6. LOG NO. DATE TIME BATTE
KM9106 Operators Manual 42 8.4.1 Setting Up Your PC A standard communications package will be able to collect the data from the Quintox. Windows Term
KM9106 Operators Manual 438.4.2 Capturing Data From Quintox To capture data direct from the Quintox into a PC the RS232 port on the Quintox has to be
KM9106 Operators Manual 44 9. CONTINUOUS MONITORING The Quintox is designed primarily as a portable emissions monitoring analyser. It can however be u
KM9106 Operators Manual 4510. Main Purge The main purge facility for Quintox is used where longer sampling of flue gas is required. It should be used
KM9106 Operators Manual 46 To access timed purge enter the following :- Enter the code 0000 at the following display :- ENTER SERVICE CODE --
KM9106 Operators Manual 47Set YES or NO depending on whether auto zero is required or not. AUTO ZERO ? NO Note: During both Manual and Timed pur
KM9106 Operators Manual 48 11. MAINTENANCE 11.1 Emptying and Cleaning the In-line Water Trap The water trap should be checked and emptied on a regula
KM9106 Operators Manual 4911.3 Charging the Battery It is important that the battery is charged on a regular basis. The instrument constantly powers
KM9106 Operators Manual 5 2. SAFETY WARNING This analyser extracts combustion gases that may be toxic in relatively low concentrations. These gases ar
KM9106 Operators Manual 50 11.5 Changing the Printer Ribbon The printer ribbon cartridge will last for approximately two rolls of paper. Remove the p
KM9106 Operators Manual 5112. PROBLEM SOLVING The following is a list of problems that may occur on the instrument through its operating life. If the
KM9106 Operators Manual 52 13. ANNUAL RE-CALIBRATION Whilst the sensors have an expected life of more than two years in normal use it is recommended t
KM9106 Operators Manual 5314. HOW TO GET EXPERT HELP There will be occasions when despite having read the manual there will be problems that you can
KM9106 Operators Manual 54 15. PRODUCT SPECIFICATION INSTRUMENT Parameter Resolution Accuracy Range Temp Measurement Flue Temperature Inlet Tempe
KM9106 Operators Manual 55HANDSET Dimensions 220 mm long 120 mm high 50 mm wide Keypad 16 tactile keys with sounder Display 4 line LCD with back
KM9106 Operators Manual 56 AMBIENT OPERATING RANGE 0 - 40°C (+32°F to 104°F) 20 - 80% RH non condensing Storage: 0-50°C Maximum gas temperature a
KM9106 Operators Manual 57APPENDICES A - Parameter Meanings The parameters and their meanings are detailed as follows : - DATE : Analyser date. See
KM9106 Operators Manual 58 NETT : Nett temperature calculated by deducting the internal AMBIENT temperature from the measured FLUE temperature. Di
KM9106 Operators Manual 59CO/CO2 R : The CO/CO2 ratio, is the ratio of measured CO divided by CO2. It gives an indication of the following :- •
KM9106 Operators Manual 6 Printed header name and telephone number 7.1
KM9106 Operators Manual 60 CxHy : Unburnt HydroCarbon reading in % of LEL of Methane, the sensor is calibrated with Methane. Displayed when a HydroCa
KM9106 Operators Manual 61B. NOx CALCULATIONS ONLY an NO Sensor fitted. working in ppm: NOx referenced to NO The user can select the assumed NO
KM9106 Operators Manual 62 NOx referenced to NO2 NOx in mg/m3 = NO in ppm multiplied by 2.05 multiplied by (1 + assumed NO2 percentage) or
KM9106 Operators Manual 63ONLY an NO2 sensor fitted When there is only an NO2 sensor fitted the NOx function is disabled NO2 in mg/m3 = NO2 in ppm m
KM9106 Operators Manual 64 C. COMBUSTION EFFICIENCY CALCULATION The efficiency calculation is based upon British Standard BS845. This identifies thr
KM9106 Operators Manual 65 Calculated data: Tnet = Net Temperature % CO2 content in flue gas % Dry Flue Gas losses % Wet losses % Unbu
KM9106 Operators Manual 66 D. CALCULATION OF FUEL DATA For any fuel not specified by Kane International the net calorific value, gross calorific valu
KM9106 Operators Manual 67E. ELECTROMAGNETIC COMPATABILITY (CE) STATEMENT This product has been tested for compliance with the following generic
KM9106 Operators Manual 74. NORMAL START UP SEQUENCE 4.1 Every Time You Use The Analyser BEFORE SWITCH-ON CHECK THAT: the Oxygen sensor is connec
KM9106 Operators Manual 8 And show the countdown screen :- -- -- -- -- -- -- -- -- -- -- -- -- -- 9 CALIBRATING 9 9 300 -- -- -- -- -- -- -- -- --
KM9106 Operators Manual 94.3 Main Displays The main display can be changed to show the following :- • Page Mode displays 4 lines of data in set for